Smartwatch ECGs read by an artificial-intelligence algorithm can flag left ventricular systolic dysfunction, but the same cutoff misclassifies patients depending on whether the tracing comes from an Apple Watch or a Samsung Galaxy Watch, researchers reported at the American Heart Association Scientific Sessions 2026.
In a study of 2211 adults who had simultaneous ECGs recorded on both devices during a clinic visit, the AI model achieved similar overall discrimination — area under the curve 0.907 for Apple Watch and 0.895 for Galaxy Watch — yet 9% of participants received discordant results. Among those positive only on the Apple Watch, 19% had LVSD; among those positive only on the Galaxy Watch, 16.7% had LVSD.
“Smartwatch AI-ECG classification depends on the device and on how repeated recordings are combined,” said Hak Seung Lee, MD, of the Digital Healthcare Institute at Sejong Medical Research Institute in Bucheon, South Korea. “Device-specific thresholds and confirmation strategies now require validation in the intended screening population, with clinical outcomes.”
A second cohort of 146 patients recorded ECGs at home twice daily for a week using mostly Galaxy Watches. Nearly half (49.3%) produced both positive and negative readings in that first week. The rate of LVSD was 5.7% with consistently negative results, 40.3% with mixed results, and 81% with consistently positive results. Requiring positive readings on at least two of three consecutive days — rather than a single positive day — cut the false-positive rate from 25.7% to 12.9% and raised the positive likelihood ratio from 3.38 to 6.35, but increased missed LVSD cases from 13.2% to 18.4%.
Averaging all first-week scores yielded a similar false-positive rate of 11.4% but missed 34.2% of LVSD cases.
“Similar discrimination across manufacturers did not ensure consistent classification,” Lee said. “At a shared cutoff, sensitivity, specificity, and individual classifications differed between Apple Watch and Galaxy Watch.”
Jessica Golbus, MD, MS, assistant professor of cardiology at the University of Michigan, called the work “an incredibly important first step” in using smartwatches to identify patients at risk for heart failure. She pointed out that the initial cohort comprised a large share of patients with established heart failure who did not have a monitoring device, and the study offered no insight into how the devices would perform across diverse patient populations spanning age, race, ethnicity, and ejection fraction.
“As the authors pointed out, we also need to understand how as a health system, we’re going to deal with a false positive artery while balancing that against the urgent and unmet need to identify individuals with asymptomatic and symptomatic LVSD,” Golbus said. “We need to ensure that the perfect strategies are in place so that when this is detected, patients can be connected with their health systems and the treatments that we know are life-changing can be instituted.”