A new federal bulletin gives stockpile managers a step-by-step way to verify whether stored N95 respirators still meet performance standards, a decision that affects hospitals, manufacturers and emergency planners weighing whether to replace aging inventory or seek emergency authorization to use it.
The National Institute for Occupational Safety and Health (NIOSH) released the Science Bulletin on Nov. 5 from its National Personal Protective Technology Laboratory. Respirator materials degrade over time, and units stored outside the manufacturer’s recommended temperature, humidity or light conditions — or held past their labeled shelf life — are no longer considered NIOSH Approved, the agency said.
The guidance matters because organizations hold millions of respirators in warehouses for years, and until now there was no standardized protocol to prove a stockpile still protects workers.
Evidence that stored units still perform can support selling inventory before it expires to fund replacements, confirming a contractor’s storage practices, or requesting emergency use authorization when demand outstrips supply — though NIOSH cautions there is no guarantee such authorization will be granted.
NIOSH recommends lot quality assurance sampling. Managers first list every respirator model and production lot, then calculate how many units each lot holds from case and box counts. Where no sampling expert is available, the agency’s stockpile study model calls for randomly selecting 40 respirators from each production lot, drawing from different pallets, cases and boxes where practical.
During collection, every pallet, case, box and unit should receive a visual check. Inspectors record signs of degradation — heavy dust, light exposure, nearby chemicals, crushed packaging or units, soiled cases, broken straps, separated nose pieces, foam sticking to neighboring respirators — to spot trends that could explain later test failures.

Third-party labs may run the tests, provided their accreditation covers all three NIOSH Standard Test Procedures for N95 filtering facepiece respirators, the bulletin states. Each test is pass/fail:
- Exhalation resistance: tested at 85 liters per minute; pass is below 25 mm water column. Minimum three units per lot.
- Inhalation resistance: tested at 85 liters per minute; pass is below 35 mm water column. Minimum three units per lot.
- Filter efficiency: pass requires no more than 5.0% particulate penetration. Minimum 40 units per lot. This test runs last because it uses a salt compound.
The same respirators can go through all three tests in sequence. NIOSH further advises testing “reference” units — newly purchased respirators of the exact same model: three for each resistance test and 20 for filter efficiency. If reference units fail, the problem may lie with the lab’s equipment or the manufacturer’s model, making it easier to attribute failures in stored units to long-term storage.

If stored respirators fail, managers should start with visual inspection notes to see whether failed units share a storage problem, and confirm temperature and humidity stayed within the manufacturer’s limits. If storage was proper and the shelf life has not expired, managers should notify NIOSH at [email protected] and the manufacturer, the bulletin advises.
When failures cluster in a single lot, pallet or case, it can indicate those respirators are no longer fit for their intended purpose, though they could be repurposed for training.
If only a few units fail with no clear pattern, NIOSH suggests testing a second sample before deciding what to do with the rest.
Even respirators that pass all three tests are not NIOSH Approved once they are past their shelf life, the authors note. The bulletin, “Monitoring Stockpiled Respirator Quality,” is available on the CDC NIOSH website.