Researchers quantitatively evaluated microstructural changes and dosimetric response of EBT-3 radiochromic films exposed to ionizing radiation in a 0.35 T magnetic field.
Because of inconsistent findings concerning dosimetric precision within magnetic fields, it is necessary to validate the use of radiochromic film dosimetry in magnetic resonance-guided linear accelerator systems. The findings indicate that magnetic fields influence topochemical polymerization directionality without affecting total polymerized material quantity.
To assess the impact of magnetic field activation, films taken from one batch received 2, 5, 8, 10, and 16 Gy doses via a 6 MV flattening filter-free photon beam delivered by a ViewRay magnetic resonance-linear accelerator unit. Scanning electron microscopy at 2500× magnification analyzed microstructural features.
Quantitative measurements of needle-like crystal structures assessed length and angular orientation using ImageJ software. Regardless of whether a magnetic field was present, crystal lengths did not change across all radiation doses, showing no statistically significant differences.
Exposure to a magnetic field caused significant angular alignment, forcing crystal structures to group within ± 80 to ± 100 degrees, contrasting with the random 360-degree distribution observed in the absence of a magnetic field. This alignment became progressively pronounced at higher doses, particularly at 8–16 Gy.
| Radiation Dose (Gy) | Crystal Length Behavior | Angular Distribution (With MF) |
|---|---|---|
| 2, 5, 8 | Consistent across doses | Clustering within ± 80 to ± 100 degrees |
| 10, 16 | Consistent across doses | Pronounced alignment clustering |
Optical density measurements employed standard RCF protocols with red-green-blue to grayscale conversion. Even with major microstructural reorientation, optical density analysis showed no meaningful dosimetric differences between magnetic field conditions, with variations staying within experimental uncertainty.
These results explain why structural changes do not compromise dose measurement accuracy. The data resolve previous conflicting literature by demonstrating that EBT-3 RCFs maintain dosimetric reliability in magnetic field environments.
The study supports continued use of these films for accurate dose measurements in magnetic resonance-guided radiotherapy applications.